Skip to main content

Labino AB

Accreditation no: 10391

Org no:
556501-1417
Phone:
08-839070
Postal and visiting address:
Fabriksvägen 9B, 186 32 Vallentuna · Show on map

Calibration, SS-EN ISO/IEC 17025:2018

Latest decision date: 2025-01-30

MätstorhetMetod/procedurProdukt/objektParameter/EgenskaperUtvidgad mätosäkerhet ±Mätområde/mätnivåFlex
IlluminanceInhouse method; TM.PM.13PhotometerWhite light3,0-3,6 %1-5000 lux
IrradiansInhouse method; TM.PM.13RadiometerBlue light4,0-5,4 %1-10000 µW/cm²
IrradiansInhouse method; TM.PM.13RadiometerUV-A4,0-4,7 %1-10000 µW/cm2
LuminanceInhouse method; TM.PM.13Luminance meterWhite light6,2-6,8 %1-3 × 10⁶ cd/m2
  • The calibration and measurement capability (CMC) includes measureand (technology area), method, measurement range and expanded measurement uncertainty. The expanded measurement uncertainty corresponds to a coverage probability (confidence level) of ~95%.
  • The accredited body has been granted flexible scope. The body shall be able to provide an updated list of the current scope of accreditation upon request.

Regulations

STAFS 2020:1
Swedacs föreskrifter och allmänna råd om ackreditering