Skip to main content

Intertechna AB

Accreditation no: 1210

Postal and visiting address:
Kvarnvägen 15, 663 40 Hammarö · Show on map

Calibration, SS-EN ISO/IEC 17025:2018

Latest decision date: 2023-05-17

MätstorhetMetod/procedurProdukt/objektParameter/EgenskaperUtvidgad mätosäkerhet ±Mätområde/mätnivå
PressureInhouse method; PGAV-1 Edition DPress showing measuring instrumentAbsolute pressure0,035 kPa + 0,0065 %0,09 kPa till 150 kPa
PressureInhouse method; PGAV-1 Edition DPress showing measuring instrumentAbsolute pressure0,050 kPa + 0,0095 %150 kPa till 200 kPa
PressureInhouse method; PGAV-1 Edition DPress showing measuring instrumentAbsolute pressure0,055 kPa + 0,0043 %70 kPa till 130 kPa
PressureInhouse method; PGAV-1 Edition DPress showing measuring instrumentAbsolute pressure0,055 kPa + 0,0095 %150 kPa till 300 kPa
PressureInhouse method; PGAV-1 Edition DPress showing measuring instrumentAbsolute pressure0,35 kPa + 0,0033 %300 kPa till 2300 kPa
PressureInhouse method; PGAV-1 Edition DPress showing measuring instrumentAbsolute pressure0,7 kPa + 0,0025 %2300 kPa till 4600 kPa
PressureInhouse method; PGAV-1 Edition DPress showing measuring instrumentAbsolute pressure0,9 kPa + 0,0029 %4600 kPa till 7000 kPa
PressureInhouse method; PGAV-1 Edition DPress showing measuring instrumentGauge pressure0,0016 kPa + 0,0175 %0,1 kPa till 5 kPa
PressureInhouse method; PGAV-1 Edition DPress showing measuring instrumentGauge pressure0,0062 kPa + 0,017 %5 kPa till 100 kPa
PressureInhouse method; PGAV-1 Edition DPress showing measuring instrumentGauge pressure0,06 kPa + 0,0036 %-100 kPa till 0,1 kPa
PressureInhouse method; PGAV-1 Edition DPress showing measuring instrumentGauge pressure0,13 kPa + 0,025 %500 kPa till 1250 kPa
PressureInhouse method; PGAV-1 Edition DPress showing measuring instrumentGauge pressure0,16 kPa + 0,0165 %100 kPa till 500 kPa
PressureInhouse method; PGAV-1 Edition DPress showing measuring instrumentGauge pressure0,35 kPa + 0,0034 %1250 kPa till 2200 kPa
PressureInhouse method; PGAV-1 Edition DPress showing measuring instrumentGauge pressure0,7 kPa + 0,0025 %2200 kPa till 4500 kPa
PressureInhouse method; PGAV-1 Edition DPress showing measuring instrumentGauge pressure0,9 kPa + 0,0021 %4500 kPa till 6900 kPa
PressureInhouse method; PGAV-1 Edition DPress showing measuring instrumentGauge pressure1,1 kPa + 0,025 %6900 kPa till 25000 kPa
  • The calibration and measurement capability (CMC) includes measureand (technology area), method, measurement range and expanded measurement uncertainty. The expanded measurement uncertainty corresponds to a coverage probability (confidence level) of ~95%.

Regulations

STAFS 2020:1
Swedacs föreskrifter och allmänna råd om ackreditering