Exceptions: Verification of deph-measurement, chapter 4.3, machine hysteresis, chapter 4,5 and Indenter, chapter 6.
Performed in the field
Hardness
Inhouse method: Shore/IRHD-HG1
IRHD gauges High
IRHD High Hardness
0,1-0,23 °IRH H
IRHD
80-100°IRH
Exceptions: Verification of the indenter, chapter 5.2.1 and pressure foot, chapter 5.2.2.
Performed in the field
Hardness
Inhouse method: Shore/IRHD-HG1
IRHD gauges Low
IRHD Low Hardness
0,1 °IRH L
IRHD
9-35°IRH
Exceptions: Verification of the indenter, chapter 5.2.1 and pressure foot, chapter 5.2.2.
Performed in the field
Hardness
Inhouse method: Shore/IRHD-HG1
IRHD gauges Micro
IRHD Micro Hardness
0,57-5,8 °IRH M
IRHD
10-100° IRH
Exceptions: Verification of the indenter, chapter 5.2.1 and pressure foot, chapter 5.2.2.
Performed in the field
Hardness
Inhouse method: Shore/IRHD-HG1
IRHD gauges Normal
IRHD Normal Hardness
0,19 - 0,33 °IRH N
IRHD
10-100° IRH
Exceptions: Verification of the indenter, chapter 5.2.1 and pressure foot, chapter 5.2.2.
Performed in the field
Hardness
Inhouse method: Shore/IRHD-HG1
Shore A gauges
Shore A Hardness
0,33 °SH
Shore
10-100 °SH
Exceptions: Verification of the indenter, chapter 5.2.1 and pressure foot, chapter 5.2.2.
Performed in the field
Hardness
Inhouse method: Shore/IRHD-HG1
Shore AM gauges
Shore AM Hardness
0,63 °SH
Shore
10-100 °SH
Exceptions: Verification of the indenter, chapter 5.2.1 and pressure foot, chapter 5.2.2.
Performed in the field
Hardness
Inhouse method: Shore/IRHD-HG1
Shore AO gauges
Shore AO Hardness
0,33 °SH
Shore
10-100 °SH
Exceptions: Verification of the indenter, chapter 5.2.1 and pressure foot, chapter 5.2.2.
Performed in the field
Hardness
Inhouse method: Shore/IRHD-HG1
Shore D gauges
Shore D Hardness
0,33 °SH
Shore
10-100 °SH
Exceptions: Verification of the indenter, chapter 5.2.1 and pressure foot, chapter 5.2.2.
Performed in the field
Mass
Inhouse method; Mass-G1
NAWI
-
0,0030 - 0,0062 mg
Mass
0,001g - 0,01g
1 measurement.
Performed in the field
Mass
Inhouse method; Mass-G1
NAWI
-
0,0030 - 0,0062 mg
Mass
0,001g - 0,01g
3 series of measurements.
Performed in the field
Mass
Inhouse method; Mass-G1
NAWI
-
0,0037 - 0,0072 mg
Mass
0,01g - 0,1g
3 series of measurements.
Performed in the field
Mass
Inhouse method; Mass-G1
NAWI
-
0,0062 - 0,01 mg
Mass
0,01g - 0,1g
1 measurement.
Performed in the field
Mass
Inhouse method; Mass-G1
NAWI
-
0,0072 - 0,014 mg
Mass
0,1g - 1g
3 series of measurements.
Performed in the field
Mass
Inhouse method; Mass-G1
NAWI
-
0,01 - 0,02 mg
Mass
0,1g - 1g
1 measurement.
Performed in the field
Mass
Inhouse method; Mass-G1
NAWI
-
0,014 - 0,031 mg
Mass
1g - 10g
3 series of measurements.
Performed in the field
Mass
Inhouse method; Mass-G1
NAWI
-
0,02 - 0,042 mg
Mass
1g - 10g
1 measurement.
Performed in the field
Mass
Inhouse method; Mass-G1
NAWI
-
0,031 - 0,13 mg
Mass
10g - 100g
3 series of measurements.
Performed in the field
Mass
Inhouse method; Mass-G1
NAWI
-
0,042 - 0,12 mg
Mass
10g - 100g
1 measurement.
Performed in the field
Mass
Inhouse method; Mass-G1
NAWI
-
0,12 - 3,4 mg
Mass
100g - 1000g
1 measurement.
Performed in the field
Mass
Inhouse method; Mass-G1
NAWI
-
0,13 - 2,5 mg
Mass
100g - 1000g
3 series of measurements.
Performed in the field
Mass
Inhouse method; Mass-G1
NAWI
-
0,37kg - 1,9kg
Mass
50kg - 500 kg
Substitution method on tank scales. On every substitution step will the resolution of the scale be added to the measurement uncetainty.
Performed in the field
Mass
Inhouse method; Mass-G1
NAWI
-
1,1kg - 1,1kg
Mass
3000kg -6000kg
Substitution method. Use of substitutions load, with an accuracy of 1/10 of the scale resolution.
Performed in the field
Mass
Inhouse method; Mass-G1
NAWI
-
1,2g - 3,3 g
Mass
50kg - 100 kg
1 measurement.
Performed in the field
Mass
Inhouse method; Mass-G1
NAWI
-
1,9kg - 7,8 kg
Mass
500kg-6000kg
Substitution method on tank scales. On every substitution step will the resolution of the scale be added to the measurement uncetainty.
Performed in the field
Mass
Inhouse method; Mass-G1
NAWI
-
2,5 - 26 mg
Mass
1000g-10 000g
3 series of measurements.
Performed in the field
Mass
Inhouse method; Mass-G1
NAWI
-
23g - 130 g
Mass
1000kg - 6000 kg
1 measurement.
Performed in the field
Mass
Inhouse method; Mass-G1
NAWI
-
26mg - 77mg
Mass
10 000g- 50 000g
3 series of measurements.
Performed in the field
Mass
Inhouse method; Mass-G1
NAWI
-
3,3g - 23 g
Mass
100kg - 1000 kg
1 measurement.
Performed in the field
Mass
Inhouse method; Mass-G1
NAWI
-
3,4 - 35 mg
Mass
1000g-10 000g
1 measurement.
Performed in the field
Mass
Inhouse method; Mass-G1
NAWI
-
35mg - 170mg
Mass
10 000g- 50 000g
1 measurement.
Performed in the field
Speed
Inhouse method; Speed-S1
Tensile tester
-
0,0048 mm/min - 5,5 mm/min
Speed
0,5-1000 mm/min
-
Performed in the field
Temperature
Measured quantity
Method / procedure
Product / object
Expanded measurement uncertainty ±
Equipment / Measurement principle
Measuring range / measuring level
Remark
Flex
Performed in the field
Temperature
Inhouse method; Temp-T1
Temperature sensors
0,21°C
Temperature
0-230°C
Bath, Ref. PRT
Performed in the field
Temperature
Inhouse method; Temp-T1
Temperature sensors
0,21°C-0,28°C
Temperature
-80°C - 0°C
Bath, Ref. PRT
Performed in the field
Temperature
Inhouse method; Temp-T1
Temperature sensors
0,29°C
Temperature
0-230°C
Block, Ref. PRT
Performed in the field
Temperature
Inhouse method; Temp-T1
Temperature sensors
0,29°C-0,34°C
Temperature
-80°C - 0°C
Block, Ref. PRT
Performed in the field
Temperature
Inhouse method; Temp-T1
Temperature sensors
0,5°C
Temperature
0-230°C
Air, Ref. PRT
Performed in the field
Temperature
Inhouse method; Temp-T1
Temperature sensors
0,50°C-0,53°C
Temperature
-80°C - 0°C
Air, Ref. PRT
Performed in the field
Temperature
Inhouse method; Temp-T1
Temperature sensors
1,8°C-1,9°C
Temperature
0°C-630°C
Bath, Ref. Thermocouples
Performed in the field
Temperature
Inhouse method; Temp-T1
Temperature sensors
1,8°C-1,9°C
Temperature
0°C-630°C
Block, Ref. Thermocouples
Performed in the field
Temperature
Inhouse method; Temp-T1
Temperature sensors
1,9°C-2,0°C
Temperature
0°C-630°C
Air, Ref. Thermocouples
Performed in the field
The calibration and measurement capability (CMC) includes measureand (technology area), method, measurement range and expanded measurement uncertainty. The expanded measurement uncertainty corresponds to a coverage probability (confidence level) of ~95%.
The accredited body has been granted flexible scope. The body shall be able to provide an updated list of the current scope of accreditation upon request.
Type 2 of flexibility:
Introduce new version of standard method and make editorial changes to non-standard method, - Introduce new version and modifications of non-standard method. The procedure must be equivalent. The changes introduced through accreditation with flexible scope must not involve new measurement principles, changes in measurement range, CMC (best measurement capability), or new quantities other than those included in the accreditation decision for calibration laboratories.