MITUTOYO Scandinavia AB
Accreditation no: 1794
- Postal and visiting address:
- Släntvägen 6, 194 61 Upplands Väsby · Show on map
Calibration, SS-EN ISO/IEC 17025:2018
Latest decision date: 2024-08-21
| Mätstorhet | Metod/procedur | Produkt/objekt | Utvidgad mätosäkerhet ± | Mätområde/mätnivå | Anmärkning | Flex |
|---|---|---|---|---|---|---|
| Length | Inhouse method; KI CMM, date 200303 Ed 1.1.0 | Coordinate measuring machine | 0,4 µm+2x10⁻⁶ L | X-axel ≤ 1,5 m | Metoden bygger på SS-EN ISO 10360-2:2002, 10360-4:2000 | |
| Length | Inhouse method; KI CMM, date 200303 Ed 1.1.0 | Coordinate measuring machine | 0,4 µm+2x10⁻⁶ L | Y-axel ≤ 3,0 m | Metoden bygger på SS-EN ISO 10360-2:2002, 10360-4:2000 | |
| Length | Inhouse method; KI CMM, date 200303 Ed 1.1.0 | Coordinate measuring machine | 0,4 µm+2x10⁻⁶ L | Z-axel ≤ 1,5 m | Metoden bygger på SS-EN ISO 10360-2:2002, 10360-4:2000 | |
| Length | Inhouse method; KI CMM, date 200303 Ed 1.1.0 | Coordinate measuring machine | 0,55 µm+1,4x10⁻⁶ L | X-axel ≤ 4,0 m | Metoden bygger på SS-EN ISO 10360-2:2009, 10360-4:2000, 10360-5:2020 | |
| Length | Inhouse method; KI CMM, date 200303 Ed 1.1.0 | Coordinate measuring machine | 0,55 µm+1,4x10⁻⁶ L | Y-axel ≤ 6,0 m | Metoden bygger på SS-EN ISO 10360-2:2009, 10360-4:2000, 10360-5:2020 | |
| Length | Inhouse method; KI CMM, date 200303 Ed 1.1.0 | Coordinate measuring machine | 0,55 µm+1,4x10⁻⁶ L | Z-axel ≤ 2,0 m | Metoden bygger på SS-EN ISO 10360-2:2009, 10360-4:2000, 10360-5:2020 | |
| Length | Inhouse method; Kl PB 0-100, date 190124 Ed 0.1.0 | Spud | 0,07 µm | 0,5 - 25 mm | - | |
| Length | Inhouse method; Kl PB 0-100, date 190124 Ed 0.1.0 | Spud | 0,08 µm | 25 - 50 mm | - | |
| Length | Inhouse method; Kl PB 0-100, date 190124 Ed 0.1.0 | Spud | 0,09 µm | 50 - 75 mm | - | |
| Length | Inhouse method; Kl PB 0-100, date 190124 Ed 0.1.0 | Spud | 0,10 µm | 75 - 100 mm | - |
- The calibration and measurement capability (CMC) includes measureand (technology area), method, measurement range and expanded measurement uncertainty. The expanded measurement uncertainty corresponds to a coverage probability (confidence level) of ~95%.
- The accredited body has been granted flexible scope. The body shall be able to provide an updated list of the current scope of accreditation upon request.
Regulations
- STAFS 2020:1
- Swedacs föreskrifter och allmänna råd om ackreditering