RTI Group AB
Accreditation no: 2021
- Postal and visiting address:
- Flöjelbergsgatan 8C, 431 37 Mölndal · Show on map
Calibration, SS-EN ISO/IEC 17025:2018
Latest decision date: 2024-12-05
| Mätstorhet | Metod/procedur | Produkt/objekt | Parameter/Egenskaper | Utvidgad mätosäkerhet ± | Mätområde/mätnivå |
|---|---|---|---|---|---|
| Current | MTB-030, utg G, 2018 | Current showing | DC | 0,1 pA, 0,1% | ±1 pA - ± 100 pA, ±100 pA - ± 10 mA |
| Current | MTB-050, utg H, 2018 | Current showing | DC | 0,13 % | 0,5 – 1500 mA |
| Voltage | MTB-010 Utg I, 2018 | Multimeter | DC | 0,56 % | 10 – 150 kV |
| Mätstorhet | Metod/procedur | Produkt/objekt | Parameter/Egenskaper | Utvidgad mätosäkerhet ± | Utrustning/Mätprincip | Mätområde/mätnivå |
|---|---|---|---|---|---|---|
| Air kerma | MTB-020, ed L, 2023 | Dosimeter | - | 1,62% | - | 18 – 150 kV |
| Air kerma | MTB-080, ed A, 2024 | Dosimeter | Half value layer (HVL) | 0,017 mm | - | 0,15 - 0,50 mm Al |
| Air kerma | MTB-080, ed A, 2024 | Dosimeter | Half value layer (HVL) | 3,5 % | - | 0,50 - 15 mm Al |
| Kerma-area-product | MTB-070, ed H, 2023 | Dosimeter | - | 3,5% | X-ray CAP | 18 – 150 kV |
| Kerma-length-product | MTB-060, ed H, 2023 | Dosimeter | - | 2% | - | 18 – 150 kV |
| Mätstorhet | Metod/procedur | Produkt/objekt | Utvidgad mätosäkerhet ± | Mätområde/mätnivå |
|---|---|---|---|---|
| Illuminance | MTB-040, ed J, 2023 | Photometer | 3,3% | 10 – 100 lux |
| Luminance | MTB-040, ed J, 2023 | Photometer | 3,3% | 10 – 1000 cd/m2 |
- The calibration and measurement capability (CMC) includes measureand (technology area), method, measurement range and expanded measurement uncertainty. The expanded measurement uncertainty corresponds to a coverage probability (confidence level) of ~95%.
Regulations
- STAFS 2020:1
- Swedacs föreskrifter och allmänna råd om ackreditering