Unfors RaySafe AB

Accreditation no: 2035

Postal and visiting address:
Björklundabacken 10, 436 57 Hovås · Show on map

Calibration, SS-EN ISO/IEC 17025:2018

Latest decision date: 2024-01-24

MätstorhetMetod/procedurProdukt/objektParameter/EgenskaperUtvidgad mätosäkerhet ±Mätområde/mätnivå
CurrentInhouse method; ACCR-1385 Utg 3Test deviceDC0,13 %0,1 - 2000 mA
Electric chargeInhouse method; ACCR-1386 Utg 4Test deviceDC0,13 %0,1 - 2000 mA
Electrical voltageInhouse method; ACCR-0454 Utg 6Test deviceNon-invasive voltage DC0,44 %18 - 40 kV
Electrical voltageInhouse method; ACCR-0454 Utg 6Test deviceNon-invasive voltage DC0,53 %40 - 155 kV
MätstorhetMetod/procedurProdukt/objektParameter/EgenskaperUtvidgad mätosäkerhet ±Mätområde/mätnivå
Air kermaInhouse method; ACCR-0453 Utg 5Test deviceAir kerma1,3 %40 - 155 kV
Air kermaInhouse method; ACCR-0453 Utg 5Test deviceAir kerma1,6 %18 - 40 kV
Air kermaInhouse method; ACCR-1112 Utg 4Test deviceAir kerma rate2,2 %18 - 155 kV
Air kermaInhouse method; ACCR-1680 edition 4Test deviceExposure time0,2 %0,1 - 100 s
Air kermaInhouse method; ACCR-1982 edition 2Test deviceHalf value layer (HVL)2,0 %40 - 155 kV, 1,4-15 mm AI
Air kermaInhouse method; ACCR-1982 edition 2Test deviceHalf value layer (HVL)2,7%18 - 40 kV, 0,2-4,0 mm AI
MätstorhetMetod/procedurProdukt/objektParameter/EgenskaperUtvidgad mätosäkerhet ±Mätområde/mätnivå
IlluminanceInhouse method; ACCR-2071 Utg 1Test deviceIlluminance2,8 %10-1000 lux
LuminanceInhouse method; ACCR-2065 Utg 1Test deviceLuminance2,7 %5-5000 cd/m2