Mätarkontroll i Stockholm AB
Accreditation no: 4409
- Org no:
- 556384-9149
- Phone:
- 08-6446870
- Postal and visiting address:
- Skarpnäcks Gårdsväg 4B, 128 31 Skarpnäck · Show on map
Calibration, SS-EN ISO/IEC 17025:2018
Latest decision date: 2021-03-31
| Mätstorhet | Metod/procedur | Produkt/objekt | Parameter/Egenskaper | Utvidgad mätosäkerhet ± | Mätområde/mätnivå | Anmärkning | Business unit |
|---|---|---|---|---|---|---|---|
| Flow | Inhouse method; FSS MO mot RM mot våg massa v2 | Mass flow meter | Fluid flow/Fluid velocity | 0,15 % | 130-280 000 kg/h | Ansl. <= 200 mm, max 65 °C, 30 – 6000 kg | |
| Flow | Inhouse method; FSS MO mot RM mot våg massa v2 | Mass flow meter | Fluid flow/Fluid velocity | 0,15 % | 6-30 000 kg/h | Ansl. <= 50 mm, max 65 °C, 3-600 kg | |
| Volume | Inhouse method; FSS MO mot RM v2 | Cold and hot water meters | - | 0,3 % | 130-280 000 kg/h | Ansl. <= 200 mm, max 65 °C, 30 - 6000 l | |
| Volume | Inhouse method; FSS MO mot RM v2 | Cold and hot water meters | - | 0,3 % | 6-30 000 kg/h | Ansl. <=50 mm, max 65 °C, 3 - 600 l | |
| Volume | Inhouse method; FSS MO mot RM mot våg v2 | Cold and hot water meters | - | 0,2 % | 130-280 000 kg/h | Ansl. <= 200 mm, max 65 °C, 100 – 6000 l | |
| Volume | Inhouse method; FSS MO mot RM mot våg v2 | Cold and hot water meters | - | 0,2 % | 6-30 000 kg/h | Ansl. <= 50 mm, max 65 °C, 5 – 600 l | |
| Volume | Inhouse method; FSS MO mot RM mot våg v2 | Cold and hot water meters | - | 0,3 % | 130-280 000 kg/h | Ansl. <= 200 mm, max 65 °C, 30 – 100 l | |
| Volume | Inhouse method; FSS MO mot RM mot våg v2 | Cold and hot water meters | - | 0,3 % | 6-30 000 kg/h | Ansl. <= 50 mm, max 65 °C, 3 – 5 l | |
| Volume | Inhouse method; SSS MO mot RM mot våg v2 | Volume meters | - | 0,3 % | 130-280 000 kg/h | Ansl. <= 200 mm, max 65 °C, 100- 6000 l | |
| Volume | Inhouse method; SSS MO mot RM mot våg v2 | Volume meters | - | 0,3 % | 6-30 000 kg/h | Ansl. <= 50 mm, max 65 °C, 10- 600 l | |
| Volume | Inhouse method; SSS MO mot RM mot våg v2 | Volume meters | - | 0,4 % | 130-280 000 kg/h | Ansl. <= 200 mm, max 65 °C, 50- 100 l | |
| Volume | Inhouse method; SSS MO mot RM mot våg v2 | Volume meters | - | 0,4 % | 6-30 000 kg/h | Ansl. <= 50 mm, max 65 °C, 5-10 l | |
| Volume | Inhouse method; SSS MO mot RM v2 | Volume meters | - | 0,4 % | 130-280 000 kg/h | Ansl. <= 200 mm, max 65 °C, 50- 6000 l | |
| Volume | Inhouse method; SSS MO mot RM v2 | Volume meters | - | 0,4 % | 6-30 000 kg/h | Ansl. <= 50 mm, max 65 °C, 5- 600 l |
- The calibration and measurement capability (CMC) includes measureand (technology area), method, measurement range and expanded measurement uncertainty. The expanded measurement uncertainty corresponds to a coverage probability (confidence level) of ~95%.
Regulations
- STAFS 2020:1
- Swedacs föreskrifter och allmänna råd om ackreditering
Inspection, SS-EN ISO/IEC 17020:2012
Latest decision date: 2022-04-13
| Type | Kravspecifikation | Kontrollobjekt | Kontrollområde | Typ av kontroll | Mätområde/mätnivå | Anmärkning | Business unit |
|---|---|---|---|---|---|---|---|
| C | STAFS 2007:2 | Heat meters | Flow sensors | Periodic inspection | Flöde 130 - 280 000 l/h, 20-6000 l Anslutn.diam. 40 - 200 mm | max temp 65°C Bänk 3 | |
| C | STAFS 2007:2 | Heat meters | Flow sensors | Periodic inspection | Flöde 5 - 30 000 l/h, 3-600 l Anslutn.diam. 15 - 50 mm | max temp 65°C Bänk 1 & 2 | |
| C | STAFS 2007:2 | Heat meters | Integrator | Periodic inspection | 100 ohm - 1000 ohm | - | |
| C | STAFS 2007:2 | Heat meters | Temperature sensors | Periodic inspection | delta T = 20, ± 2 °C | - | |
| C | STAFS 2007:2 | Heat meters | Temperature sensors | Periodic inspection | delta T = 30, ± 2 °C | - | |
| C | STAFS 2007:2 | Heat meters | Temperature sensors | Periodic inspection | delta T = 50, ± 2 °C | - | |
| C | STAFS 2007:2 | Volume meter | Cold water | Periodic inspection | Flöde 130 - 280 000 l/h, 20-6000 l Anslutn.diam. 40 - 200 mm | max temp 30 °C Bänk 3 | |
| C | STAFS 2007:2 | Volume meter | Cold water | Periodic inspection | Flöde 5 - 30 000 l/h, 2-600 l Anslutn.diam. 15 - 50 mm | max temp 30 °C Bänk 1 & 2 | |
| C | STAFS 2007:2 | Volume meter | Hot water | Periodic inspection | Flöde 130 - 280 000 l/h, 20-6000 l Anslutn.diam. 40 - 200 mm | max temp 65 °C Bänk 3 | |
| C | STAFS 2007:2 | Volume meter | Hot water | Periodic inspection | Flöde 5 - 30 000 l/h, 3-600 l Anslutn.diam. 15 - 50 mm | max temp 65 °C Bänk 1 & 2 |
Regulations
- ILAC G17:01/2021
- ILAC Guidelines for Measurement Uncertainty in Testing
- STAFS 2020:1
- Swedacs föreskrifter och allmänna råd om ackreditering